3.4 Starting a Test
______________________________________________________________________________________________
"SETTING" State
Key-lock Function
Optional Function
READY
態」での設
Starting a Test
Press the
START
key when
READY
is lit. The unit will change to the TEST status and a test
will start.
TEST
and the
DANGER
lamp are lit in the TEST state.
If a comparative-voltage value has been set, the test will not start until the output voltage is
within
5% of the comparative-voltage value (output voltage 1kV 50 V).
READY
態」での設
Determination
PASS
/
FAIL
screening is conducted based on whether a measured-current value exceeds the
upper- or lower-limit value. The test is failed (FAIL) if, with a comparative-voltage value set,
the output-voltage value fails to reach the comparative-voltage range within 5 seconds after
the
START
key is pressed, or if the output-voltage value deviates from the comparativevoltage
range.
See Section 3.3
See Section 3.2.1
See Cha
p
ter 6
See Section 3.4.2
See Section 3.5
READY
state
Setting the test parameters
READY
態」での設
Setting the "Output voltage"
Before starting a test, make output voltage settings using the output-voltage knob.
A voltage is output during output-voltage setting using the output-voltage knob. Never touch
the HIGH terminal, H.V. TEST LEAD, or tested object.
See Section 3.4.1
3.4 Starting a Test
The flowchart below explains how a test is carried out.