131
_____________________________________________________________________________________________
9.4 Command
______________________________________________________________________________________________
9.4.2 Specific Command Messages
Syntax
:WITH:VREF <data>
<data>
OFF/ON
Response
syntax
OK
Voltage Comparator ON/OFF settings completion.
Function
Enables and disables the Voltage Comparator for withstand voltage tests, in the READY state.
Example
Enables the Voltage Comparator for a withstand voltage test.
Transmission :WITH:VREF ON
Voltage Comparator enablement is ON.
Response OK
Settings completion
Error
The execution of this command in a state other than the READY state causes an execution error.
Syntax
:WITH:VREF?
Response
syntax
<data>
<data>
OFF/ON
Function
Queries the Voltage Comparator enablement for withstand voltage tests.
Example
Queries the Voltage Comparator enablement for withstand voltage tests.
Transmission :WITH:VREF?
Queries the Voltage Comparator enablement.
Response ON
The Voltage Comparator enablement is ON.
Syntax
:WITH:CLOW <data>
<data>
OFF/ON
Response
syntax
OK
Withstand voltage test lower-limit value ON/OFF settings completion
Function
Enables and disables the lower-limit value for withstand voltage tests, in the READY state.
Example
Enables the lower-limit value for a withstand voltage test.
Transmission :WITH:CLOW OFF
The withstand voltage test lower-limit value enablement is ON.
Response OK
Settings completion
Error
The execution of this command in a state other than the READY and SETTING states causes an
execution error.
Note that an execution error occurs when setting the test lower limit to on when the test lower
limit is equal to or greater than the test upper limit.
:WITH:VREF
Enables and disables the Voltage Comparator for withstand voltage tests.
:WITH:VREF?
Queries the Voltage Comparator enablement for withstand voltage tests.
:WITH:CLOW
Enables and disables the lower-limit value for withstand voltage tests.