5.5 PASS or FAIL Determination
______________________________________________________________________________________________
5.5.1 "PASS" State
WARNING
Even when a test has been terminated, there may still be voltage in the
output-voltage terminal. Before touching the output-voltage terminal,
H.V. TEST LEAD, or tested object, confirm the following items.
(1) The analog voltmeter reads 0 kV.
(2) The
DANGER
lamp is OFF.
(3) The
READY
lamp is lit (it is off in the Double Action mode).
5.5 PASS or FAIL Determination
When the preset test time has elapsed, the unit switches to the PASS state
and immediately stops outputting a voltage.
To the test, press the STOP key, which will forcibly terminate the test.
Optional Functions
The PASS state is held using the PASS Hold function. (See Section 6.1)
If the W or I key is pressed when the Hold function is enabled, you can
view the next test result.