5.4 Starting a Test
______________________________________________________________________________________________
Setting the test parameters
Select W I mode, or I Wmode.
「
READY
状態」での設定
Starting a Test
Press the
START
key when
READY
is lit. The unit will change to the TEST status and a test
will start.
TEST
and the
DANGER
lamp are lit in the TEST state.
「
READY
状態」での設定
Determination
Compares the preset test settings (current upper- and lower-limit values, and resistance
upperand
lower-limit values) and the measured value, and conducts screening. Furthermore, when
performing a withstand voltage test and the output-voltage value deviates from the
comparative-voltage range, the test results in a FAIL.
See Section 5.4.1
See Section 5.5
Make withstand-voltage test settings in withstand-voltage mode. (See chapter 3.)
Make insulation-resistance test settings in insulation-resistance mode. (See chapter 4.)
Selecting the W I/I Wmode
5.4 Starting a Test
The flowchart below explains how a test is carried out.