132
_____________________________________________________________________________________________
9.4 Command
______________________________________________________________________________________________
Syntax
:WITH:CLOW?
Response
syntax
<data>
<data>
OFF/ON
Function
Queries the lower-limit value enablement for withstand voltage tests.
Example
Queries the lower-limit value enablement for withstand voltage tests.
Transmission :WITH:CLOW?
Queries the lower-limit value enablement for withstand voltage tests.
Response OFF
The lower-limit value enablement is OFF.
Syntax
:WITH:TIM <data>
<data>
OFF/ON
Response
syntax
OK
Withstand voltage test time ON/OFF settings completion
Function
Enables and disables the test time for withstand voltage tests, in the READY state.
Example
Enables the test time for a withstand voltage test.
Transmission :WITH:TIM ON
The withstand voltage test time enablement is ON.
Response OK
Settings completion
Error
The execution of this command in a state other than the READY state causes an execution error.
Syntax
:WITH:TIM?
Response
syntax
<data>
<data>
OFF/ON
Function
Queries the test time enablement for withstand voltage tests.
Example
Queries the test time enablement for withstand voltage tests.
Transmission :WITH:TIM?
Queries the withstand voltage test time enablement.
Response ON
The test time enablement is ON.
:WITH:CLOW?
Queries the lower-limit value enablement for withstand voltage tests.
:WITH:TIM
Enables and disables the test time for withstand voltage tests.
:WITH:TIM?
Queries the test time enablement for withstand voltage tests.