53
_____________________________________________________________________________________________
4.3 "SETTING " State
______________________________________________________________________________________________
Setting range (M
Ω
) Setting resolution (M
Ω
)
0.20 to 2.00 0.01
2.10 to 20.0 0.1
21.0 to 200 1
210 to 2000 10
Test voltage
(V)
Lower-value
(M
Ω
)
Range
Measurement
range (M
Ω
)
500 0.20 to 2.00 2M
Ω
0.17 to 10.0
*1
2.10 to 20.0 20 M
Ω
1.70 to 100
21.0 to 200 200 M
Ω
17.0 to 1000
210 to 2000 2000 M
Ω
170 to 2200
*2
1000 0.20 to 4.00 4M
Ω
0.01 to 20.0
*3
4.10 to 40.0 40 M
Ω
3.40 to 200
41.0 to 400 400 M
Ω
34.0 to 2000
410 to 2000 2000 M
Ω
340 to 2200
*2
Upper- or lower-value
Items related to test voltages, lower-limit values and ranges, and
measurement ranges
*1 The range from 0.17 to 0.49 is outside the guaranteed range of accuracy.
*2 The range from 2001 to 2200 is outside the guaranteed range of
accuracy.
*3 The range from 0.01 to 0.99 is outside the guaranteed range of accuracy.
When the resistance value exceeds the measurement range, it is displayed as
"O.F." (overflow).
When the set resistance value does not reach the measurement range, values
are displayed on the "U.F." (underflow).
Optional Functions
You can make the low measurement range a fixed range or an automatic
range, depending on the optional function settings. (See Section 6.10)