Momentary-out
The momentary OUT function allows current output only while the
START
key
is held down.
The
START
key on the remote control or the START signal via external I/O
has the same effect.
Selection
0
: Not set (initial setting)
1
:Set
5
Double action
Enable this function to allow testing to start only when the
START
key is
pressed within about 0.5 seconds after the
STOP
key.
Selection
0
: Not set (initial setting)
1
:Set
6
FAIL mode
Enable this function to restrict hold release to the
STOP
key on the main unit.
Selection
0
: Not set (initial setting)
1
:Set
7
RS Command [START]
Turn this function on to enable the RS-232C START command.
Selection
0
: Not set (initial setting)
1
:Set
8
Inter-lock function
Enable this function to activate the external I/O interlock terminals.
Selection
0
: Not set (initial setting)
1
:Set
9
Voltage Comparator Position
When the voltage comparator is set to ON for a withstand voltage test, select
whether you want to use the voltage comparator when starting and during the
test, or only when terminating the test.
Selection
0
: Start test (initial setting)
1
:Endtest
10
Insulation Resistance Test measurement range
Select whether you want to use a fixed or an automatic range as the insulation
resistance test measurement range. The fixed ranges are automatically selected
depending on the preset lower-limit value. Auto range switches between ranges
depending on the measured value, but it takes time to display this value, since
it is displayed after the range is switched. (This takes approximately 1.3
seconds.)
Selection
0
: Fixed range (initial setting)
1
: Auto range
11
Insulation Resistance Test Termination Mode
When performing an insulation resistance test, set whether you want to conduct
the test for the set test time regardless of the decision, terminate the test when
PASS screening is performed, or terminate the test when FAIL screening is
performed.
This mode is effective when the test time is set.
Selection
0
: Test for set time (initial setting)
1
: Terminate test at PASS screening
2
: Terminate test at FAIL screening