116
────────────────────────────────────────────────────
8.3.6 Specific Commands for Others
────────────────────────────────────────────────────
Syntax
:ESE0?
Response
syntax
(Header ON)
ESE0 <0-255>
(Header OFF)
<0-255>
Example
Transmission
Response
:ESE0?
(Header ON)
:ESE0 128
(Header OFF)
128
Syntax
:ESR0?
Response
syntax
(Header ON)
ESR0 <0-255>
(Header OFF)
<0-255>
Example
Transmission
Response
:ESR0?
(Header ON)
:ESR0 128
(Header OFF)
128
Terminates the reading out of the file.
Syntax
:COM:TERMinator <0/1/2>
<0>: LF
<1>: CR+LF
<2>: CR (when powering on)
Example
Transmission
:COM:TERM 1
Set the terminator of response message
to CR+LF.
Function
Returns the contents of the event status
enable register 0 (ESER0) as a NR1
value (0 to 255).
128 64 32 16 8 4 2 1
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
EOFL EOWL EOSL EORO Reserved
Notes
With this query, if any error occurs, no
response message is produced.
Function
Returns the contents of SESR as NR1
value (0 to 255), and clears the data.
128 64 32 16 8 4 2 1
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
EOFL EOWL EOSL EORO Reserved
Notes
With this query, if any error occurs, no
response message is produced.
Function
Sets the response message terminator
of RS-232C inerface.
:ESE0? (GP-IB only)
Queries the event status enable register 0.
:ESR0?
Queries the event status register.
:COM:TERMinator (RS-232C only)
Sets the terminator of response message.