131
────────────────────────────────────────────────────
10.2 Function Specifications
────────────────────────────────────────────────────
Maximum output voltage
Maximum 10 V (RL= )
Amplitude setting range
10 V range 0 to 10 V, open-circuited (resolution 1 mV)
1 V range 0 to 1 V, open-circuited (resolution 0.1 mV)
0.1 V range 0 to 0.1 V, open-circuited (resolution 0.01 mV)
DC offset setting range
10 V range -10 to 10 V, open-circuited (resolution 1 mV)
1 V range -1 to 1 V, open-circuited (resolution 0.1 mV)
0.1 V range -0.1 to 0.1 V, open-circuited (resolution 0.01 mV)
Minimum load impedance
40 Ω
Output impedance
50 Ω 2% (DC)
Square wave rise and fall times
Within 45 ns (10 to 90% of maximum amplitude, when the setting of
LPF "pass", RL=50 Ω)
Square wave overshoot
Less than 5% of set amplitude (P-P value, with LPF bypassed, RL=50
Ω
Inter-Channel Skew
Within 25 ns (measured with the same waveform selected)
Except during sweeping the frequency.
Output range accuracy
1 V range Add 0.2% of the range to the accuracy of the 10 V range
0.1 V range Add 0.4% of the range to the accuracy of the 10 V range
For specification of 10 V, see
FG Function, and Arbitrary
waveform generator function.
Type of waveform
Sine, square (constant duty), triangle, ramp-up, ramp-down and pulse
waveforms, noise and DC
Frequency range
Sine, square: 0 to 10 MHz (resolution 10 mHz)
Triangle, ramp, pulse: 0 to 200 kHz (resolution 10 mHz)
Frequency accuracy
Within (50 ppm of setting value + 50 μHz)
DC offset accuracy
Within (0.5% of setting value + 25 mV)
DC offset stability
Within (DC offset accuracy X 0.1)/
Amplitude accuracy
Within (2% of setting value + 20 mVrms) at 1 kHz, sine wave
Amplitude stability
(amplitude accuracy X 0.1)/
Phase difference
-360.00 to 360.00 deg (resolution 0.01 deg)
Jitter (triangle, ramp and pulse
waveforms)
Within 100 ns p-p
Square wave duty cycle
40 to 60% fixed
Pulse wave duty cycle
adjustable range
1 to 99% (setting resolution: the larger of 0.1% or 100 ns) valid for
pulse widths greater than 100 ns
10.2 Function Specifications
Analog Output (common to both function generator and arbitrary waveform outputs)
FG Function (Accuracy measured in the 10 V range)