2 - 6
(cont’d)
Alarm
Control
No.
Alarm Category
Sub-
code
Level Description Remedy
101 to
106
SAMPLE
DISK
8 5 STOP At resetting, the sample disk
cannot detect the home
position.
Check the home
detector.
8 6 STOP At resetting, the sample disk
does not stop at the specified
point of home position.
Same as above
107 SAMPLE
DISK
8 7 STOP At resetting, the sample disk
does not leave the home
position.
Check the home
detector.
251 to
310
SAMPLE
SHORT
11 1 to 60WARNING In sipping from the sample
cup, sample on the sample
disk is inadequate.
Sub-code groups
Routine/
stat samples:1 to 35
Control STD:36 to 57
W1 to W3 :58 to 60
Add sample.
551 to
552
SAMPLE
SYRINGE
15 1 S.STOP/
STOP
The serum syringe does not
reach the upper dead point.
Check the upper
dead point detector.
15 2 S.STOP/
STOP
The serum syringe does not
go down from the upper dead
point.
Same as above
561 to
568
REAGENT
PROBE
16 1 STOP The reagent probe does not
reach the upper dead point in
ascending motion.
Check the upper
dead point detector.
16 2 STOP The reagent probe moves
down abnormally in
descending motion.
Check liquid level
detection.
16 3 STOP The reagent probe does not
go down from the upper dead
point in descending motion.
Check the upper
dead point detector.
16 4 WARNINGDetection of abnormal
descending motion of the
reagent probe remains on.
Check the
descending error
detector.