2 - 7
(cont’d)
Alarm
Control
No.
Alarm Category
Sub-
code
Level Description Remedy
561 to
568
REAGENT
PROBE
16 5 STOP When the reagent probe
turns toward the cell, it
cannot detect the cell
position.
Check the home
detector.
16 6 STOP When the reagent probe
turns from the cell side to
other position, it does not
leave the cell position.
Same as above
16 7 STOP Before the probe goes down,
the liquid level detector is
already turned on.
Check liquid level
detection.
16 8 STOP In rotation of the reagent
probe, it is not set at the
upper dead point.
Check the upper
dead point detector.
581 to
593
REAGENT
DISK
18 1 STOP The stop position of reagent
disk cannot be detected.
Check home
detection or count
detector.
18 2 STOP The reagent disk does not
stop at the specified position.
Same as above
18 3 STOP The home position of reagent
disk cannot be detected.
Check home
detection.
621 to
622
REAGENT
SYRINGE
22 1 STOP The reagent syringe does not
reach the upper dead point.
Check the upper
dead point detector.
22 2 STOP The reagent syringe does not
move down from the upper
dead point.
Same as above
641 ISE
SIPPER
24 1 STOP The sipper nozzle does not
reach the upper dead point
(during resetting/
operation).
Check the upper
dead point detector of
ISE sipper.
642 ISE
SIPPER
24 2 WARNING/
STOP
(Note 4)
The sipper nozzle does not
leave the upper dead point.
Same as above
661 to
662
ISE
SYRNG
26 1 WARNING/
STOP
(Note 4)
The sipper syringe does not
reach the upper dead point.
Check the upper
dead point detector of
ISE Syringe.
26 2 WARNING/
STOP
(Note 4)
The sipper syringe does not
leave the upper dead point.
671 ISE STOP
OK ?
27 1 WARNING/
STOP
(Note 4)
The ISE function is stopped
due to alarm.
(This warning is indicated
when restart in the sampling
stop status was attempted.)
Check alarm log and
deal with the alarm.
NOTES: 3. S.STOP may be issued only during operation.
4. Though the photometry assay function works, the ISE function does not work.