2 - 13
Alarm
Control
No.
Alarm Category
Sub-
code
Level Description Remedy
3101 to
3189
STANDARD? 83 1 to 40
(ch.)
WARNING <<Photometry assay>>
(1) In calibration, the
STD absorbance data
is indicated with
alarm.
(2) In calibration, data
calculation is
disabled.
<<ISE assay>>
(1) In calibration, the
potential data of
standard or internal
standard solution is
indicated with ADC
error, insufficient
sample alarm, noise
alarm or level alarm.
(2) In calibration, data
calculation is
disabled.
(The result of calibration
is not updated nor saved
onto FD.)
(Photometry assay)
•Replace STD
sample.
•Check the
concentration
parameter.
(ISE assay)
Same as above
3251 to
3291
CALIBRA-
TION
84 1 to 40
(ch.)
WARNING <<Photometry assay>>
The factor value 'K'
determined in calibration
differs from the previous
value by more than ±20%.
<<ISE assay>>
The calibrator
concentration value and
slope value determined in
calibration differ from the
previous values by more
than the compensation
limit (%).
(Photometry assay)
(ISE assay)
Same as above
3401 to
3438
CALIB.SD? 85 1 to 36
(ch.)
WARNING The mean error determined
in multi-point calibration is
larger than the SD limit
(input value).
Replace STD and
check again.
(cont’d)