(b) Transfer of absorbance data in entire reaction process (from analyzer to host)
1) Specification of size
The size of text is specifiable on the communication parameter screen.
Select either 256 or 512 bytes. On selection of 256 or 512 bytes, text is
transferred divided as shown below.
(i) When 256-byte mode is specified for text size
1B 1B 2B 37B 10B × 4
First STX 1 Fu Sample information Analytical data 1
6B × 4 3B 6B × point count
Analytical data 4 BLANK1 BLANK4 Point count ABS1
1B 1B
ABS24 ETX BCC (Variable)
1B 1B 2B 37B 3B 6B × point count
Final STX : Fu Sample information Point count ABS25
ABS35 ETX BCC
(ii) When 512-byte mode is specified for text size
1B 1B 2B 37B 10B × 4
Final STX : Fu Sample information Analytical data 1
6B × 4 3B 6B × point count
Analytical data 4 BLANK1 BLANK4 Point count ABS1
1B 1B
ABS35 ETX BCC (Variable)
2) Transfer unit
This text is transferred in units of channel. Even when the text size is 256
bytes, transfer is completed in a single text if the point count is 24 or less.
The frame character at that time is not '1' but ':'.
3) Sample information
Refer to "sample information" in (2) of 16.6.2.