16 - 34
4) Analytical data 1 to 4 (10 characters each)
(i) For transfer format, refer to (2) of 16.6.1.
(ii) Table 16-18 is followed when there is no relevant test for analytical data
1 to 4.
(iii) When two-channel simultaneous measurement is specified, data for two
channels is transferred, and data for up to four channels (1 channel + L,
H, I) is transferred when serum index measurement is specified.
Table 16-18
Setting
Channel No. " bbb "
Measured value " bbbbbb "
Data alarm " b "
5) BLANK 1 to 4 (6 characters each)
The transfer format of each cell blank data is shown below.
6) Point count (3 characters)
The number of photometric points to be transferred in one text is transferred.
Table 16-19
Reaction Time 3 min 4 min 5 min 10 min
Point Count 11 14 17 35
7) ABS 1 to 35
Absorbance data in the entire reaction process (difference data between two
wavelengths at each photometric point) is transferred in the same format as of
the above cell blank data. When the point count is less than 35, data is
closely transferred in sequence starting from ABS 1.
10