7 REFERENCE CONFIG MENU
67
7.5.3 Selecting the Relock Speed (Manual)
When RECOVERY MODE is set to MANUAL, to select the relock speed, follow the procedure
below.
From the time this instrument is started until the time set with IMMEDIATE MODE TIME elapses,
the instrument operates with IMMEDIATE regardless of the selection made here.
Procedure
REFERENCE CONFIG > RECOVERY/TRACKING > MANUAL SETTING
Parameters
IMMEDIATE: Immediately relocks onto the reference signal.
FAST: Quickly relocks onto the reference signal.
SLOW: Slowly relocks onto the reference signal.
Initial value
IMMEDIATE
7.5.4 Power-on Settings
From the time this instrument is started until the set time elapses, the instrument operates with
IMMEDIATE regardless of the selection made with AUTO SETTING or MANUAL SETTING.
To set the time for which the instrument operates with IMMEDIATE, follow the procedure below.
Procedure
REFERENCE CONFIG > RECOVERY/TRACKING > IMMEDIATE MODE TIME
Parameters
OFF / 5 - 30 [min]
Initial value
5 [min]
7.6 Setting the Relock
When REFERENCE SOURCE is set to an option other than INTERNAL, relock onto the reference
signal, select OK by following the procedure below.
Procedure
REFERENCE CONFIG > REFERENCE READJUST