EasyManuals Logo

Micromeritics ACCUPYC II 1345 User Manual

Default Icon
239 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #164 background imageLoading...
Page #164 background image
12 Maintenance and Troubleshooting
Action B: Test the outlet. If there is no power, contact electrician.
Cause C: Plug prongs bent so that contact not made at outlet.
Action C: Unplug the power cable and re-plug into outlet. If indicator does not illuminate,
have electrician adjust prongs or replace outlet or plug.
Cause D: Power cord damaged.
Action D: Have electrician check cord using test meter. Replace if defective.
Cause E: Loose internal connection, broken wire, or failure of internal power supply.
Action E: Contact a Micromeritics Service Representative for repair or replacement
information.
Specified pressure not reached or maintained.
Cause A: Chamber cap not properly closed.
Action A: Close chamber cap by turning fully clockwise.
Cause B: Chamber cap contains dust or debris or the O-ring is not properly greased.
Action B: Using a lint free tissue, clean the chamber cap and the rim of the cell chamber.
Lightly grease the chamber cap O-ring. See Chamber Cap O-Ring on
page12 - 5.
Cause C: The chamber cap O-ring is not properly seated.
Action C: Check the chamber cap. Ensure that the O-ring is properly seated and that it
contains no scratches or cuts.
Cause D: The chamber cap O-ring is cut or scratched.
Action D: Replace the O-ring in the chamber cap. See Chamber Cap O-Ring on
page12 - 5.
Cause E: Gas leaks in the cell chamber or expansion chamber.
Action E: Check the pycnometer for leaks. SeeCheck the Cell and Expansion Cham-
bers for Leaks on page12 - 7.
Cause F: The helium tank is low on gas or empty. Tank pressure should be at least 200
psi above regulator pressure.
Action F: Check tank. Minimum recommended pressure is 200 psig above regulator pres-
sure. See Regulator Pressure on page12 - 15.
Cause G: The shut-off valve on the gas cylinder is closed.
Action G: Ensure that the valve is open.
Cause H: The Zero offset (of pressure transducer) is too low.
12 - 2
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) Jan 2021

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ACCUPYC II 1345 and is the answer not in the manual?

Micromeritics ACCUPYC II 1345 Specifications

General IconGeneral
BrandMicromeritics
ModelACCUPYC II 1345
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals