EasyManua.ls Logo

Micromeritics ACCUPYC II 1345 - Calibrate Using the Keypad; Calibrate Function

Default Icon
239 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 About Calibration
CALIBRATE USING THE KEYPAD
n Calibrate Using the Software on page10 - 1
Use these instructions for keypad installations only. If a computer is attached to the
module, skip this section. All setup functions will be done in the analysis software.
The pycnometer volume should be calibrated every time it is restarted. Cell volume and expansion
volume in the set up parameters are updated automatically upon calibration. Run an analysis with an
empty cup to see how close the average volume is to 0. It should be ±0.05% of full scale. If it is not
within ±0.05% of full scale, calibration is required.
CALIBRATE FUNCTION
This command performs a full volume slope and offset calibration. Press ALT + . (decimal) to
access the Calibrate function.
Display Text
Description
Calibration Type?
Volume
Press CHOICE until the option is displayed, then press ENTER.
n Volume
n Temperature
n Pressure
n Copy to an external device
n Load from an external device
n Reset Pressure Calibration
10 - 14
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) Jan 2021

Table of Contents

Related product manuals