10 About Calibration
REVIEW CALIBRATION
The Review function reviews the results of the last calibration operation.
The sample cell volume and the expansion cell volume are used for calculating the sample volume.
The cell volume and expansion volume are updated automatically when the pycnometer is
calibrated.
The prompts that display during a review of calibration data depend on the options specified inSetup
> Analysis Parameters and Report Options.
1. Press Alt + 5 to display the Chamber Insert prompt from which the insert can be selected (if
used).
2. Press ENTER to display the volume of the calibration standard. This volume can be edited
from this prompt.
3. Press ENTER to display the Which Chamber prompt.
4. Press CHOICE to select whether to view the results for the sample chamber or the expansion
chamber.
5. Press ENTER to display a prompt showing the starting time and date.
6. Press ENTER to display a prompt showing the ending time and date.
7. Press ENTER to display the first of the prompts containing data. The [n] in this example rep-
resents the cycle number.
8. Continue pressing ENTER to view all of the data. Press CHOICE to exclude data from report
calculations. An asterisk will display indicating that it will be excluded.
9. Press SAVE to return to the Reload prompt.
10. Press Alt + 6 to print report results.
Display Text
Description
Chamber insert?
(insert)
Displays the insert selected for the calibration or None if an insert was
not used. If the selection is incorrect, press CHOICE until the correct
one is shown, then press ENTER.
n None, 0.1 cm
3
(for 1 cm
3
unit)
n None, 3.5 cm
3
, 1.0 cm
3
(for 10 cm
3
unit)
n None, 35 cm
3
, 10.0 cm
3
(for 100 cm
3
unit)
n None, 650 cm
3
, 1300 cm
3
(for 2000 cm
3
pycnometer)
10 - 18
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) — Jan 2021