HANDLING SYSTEM COMPONENTS
CALIBRATION STANDARD
n Wear latex or nitrile, powder-free gloves to prevent transfer of oil from hands.
n Do not drop the standard into the sample cup. Roll the standard into the cup to prevent damage.
n Always return the standards to the case. Standards are unit specific.
SAMPLE CHAMBER CAP
n Wear latex or nitrile, powder-free gloves to prevent transfer of oil from hands.
n Keep the cap on the sample chamber except when inserting or removing the sample cup.
When left uncapped, the sample chamber temperature may become unstable and/or
water vapor will adsorb on the inner surface of the chamber. Either of these conditions
can affect analysis results.
n Avoid laying the chamber cap on a work surface. Debris may collect on the greased surface of the
O-ring which can also affect analysis results.
n When multiple analysis modules are connected, never interchange sample chamber caps. Caps
are unit specific.
Handling System Components
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) Jan 2021
11 - 5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ACCUPYC II 1345 and is the answer not in the manual?

Micromeritics ACCUPYC II 1345 Specifications

General IconGeneral
BrandMicromeritics
ModelACCUPYC II 1345
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals