HANDLING SYSTEM COMPONENTS
CALIBRATION STANDARD
n Wear latex or nitrile, powder-free gloves to prevent transfer of oil from hands.
n Do not drop the standard into the sample cup. Roll the standard into the cup to prevent damage.
n Always return the standards to the case. Standards are unit specific.
SAMPLE CHAMBER CAP
n Wear latex or nitrile, powder-free gloves to prevent transfer of oil from hands.
n Keep the cap on the sample chamber except when inserting or removing the sample cup.
When left uncapped, the sample chamber temperature may become unstable and/or
water vapor will adsorb on the inner surface of the chamber. Either of these conditions
can affect analysis results.
n Avoid laying the chamber cap on a work surface. Debris may collect on the greased surface of the
O-ring which can also affect analysis results.
n When multiple analysis modules are connected, never interchange sample chamber caps. Caps
are unit specific.