REVIEW ANALYSIS
Use these instructions for keypad installations only. If a computer is attached to the
module, skip this section. All setup functions will be done in the analysis software.
The Review function allows review of the results of the last analysis operation. Press Alt + 5 to
review or edit analysis results. The analysis information can also be edited through the prompts. The
prompts that display during the analysis review depend on the options selected in Setup > Analysis
Parameters and Report Options.
1. Press Alt + 5 to display the Sample ID prompt containing the sample identification.
2. Press ENTER to display the Description prompt containing an additional description. This
prompt typically is not used unless a keyboard is connected to the analyzer.
3. Press ENTER to display the Sample Mass prompt containing the sample’s mass.
4. Press ENTER. The Start prompt containing the start date and time of the analysis is dis-
played.
5. Press ENTER. The End prompt containing the ending date and time of the analysis is dis-
played.
6. Press ENTER. The first of the prompts containing data is displayed. The [n] displayed in the
example prompts represents the cycle number.
7. Continue pressing ENTER to view data. Press CHOICE to exclude data from report cal-
culations. An asterisk will display indicating that it will be excluded. Press CHOICE again to
remove the asterisk.
8. Press SAVE to return to the Reload prompt.
9. Press Alt + 6 to print report results.
Display Text
Description
Sample ID:
(sample identification)
Identification entered for the analysis.
Sample Mass:
(mass)
Sample mass for the analysis (if Density was chosen as Analysis
display mode).
Temperature:
(sample chamber temp)
Sample chamber temperature.
Start:
(time) (date)
Time and date the analysis began.
Time: HH:MM:SS
Date: DD/MM/YY
Review Analysis
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) — Jan 2021
6 - 9