CALIBRATE TEMPERATURE OFFSET
Unit [n] >Calibration >Temperature
Opens the Calibrate Temperature Offset box.
Calibrates the temperature offset. Use an external reference device to determine the sample cell
temperature. Enter the value in the field, then click Done.
A Calibration report is not generated for the temperature offset. The entry is recorded in the
Instrument Log. To view the log, go to Unit [n] > Show Instrument Log.
1. Enter the current sample cell temperature.
2. Click Done. Temperature reading will be adjusted to the entered temperature.
Calibrate Temperature Offset
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) — Jan 2021
10 - 3