Do not expose the system to excessive vibration through
transportation or irregular hard / soft surfaces when moving.
Mechanical (casters) and electronic (monitor) damage may
result. If the system is moved or transported often in
questionable circumstances, contact Mindray Customer
Service Department or sales representative.
5
Do not connect this system to outlets with the same circuit
breakers and fuses that control the current of devices such
as life-support systems. If this system malfunctions and
generates overcurrent, or when there is an instantaneous
current at power ON, the circuit breakers and fuses of the
building’s supply circuit may be tripped.
6
Always keep the system dry. Avoid transporting this system
quickly from a cold place to a warm place; otherwise
condensation or water droplets may form allowing a short
circuit and possible electric shock.
7
If the circuit breaker is tripped, it indicates that the system or
a peripheral device was improperly shut down and the
system is unstable. You cannot repair the system under this
circumstance and must call the Mindray Sales or Customer
Service Representative.
8
There is no risk of high-temperature burns during routine
ultrasound examinations. It is possible for the surface
temperature of the transducer to exceed the body
temperature of a patient due to environmental temperature
and exam type combinations. If a patient complains of any
excessive heat from the transducer, immediately stop
scanning. To prevent patient burns, ensure there is no
surface damage to the transducer. Do not apply the
transducer to the same region on the patient for a long time.
Apply the transducer only for a period of time required for
the purpose of diagnosis.
9
The system and its accessories are not disinfected or
sterilized prior to delivery. The operator is responsible for the
cleaning and disinfection of transducers and sterilization of
biopsy brackets according to the manual, prior to the use.
All items must be thoroughly processed to completely
remove harmful residual chemicals or gasses that are
harmful to the human body or that may damage the
accessory.
10
It is necessary to press [End Exam] to end the current scan
that is in progress and clear the current Patient Information
field. Otherwise new patient data may be combined with the
previous patient data.