Probes and Biopsy
18-36
(5) Insert a biopsy needle with the same specification as that of the guiding block
into the hole of the guiding block.
îš„ NGB-007and NGB-012 (taking NGB-007 as example)
Metal/needle detachable needle-guided bracket:
(1) Put on the sterile probe sheath.
(2) Hold the probe by one hand, select the proper needle-guided bracket, and hold it
with the other hand. Match the groove and tab with the tab and groove of the
probe respectively. Amount the bracket onto the probe.
(3) Screw the pinch nut of the needle-guided bracket to confirm that the
needle-guided bracket is properly installed on the probe.
(4) Select a proper guiding block and push it into the groove above the angle block,
and clamp it tightly.
(5) Screw the nut of the block to secure the block.
(6) Insert a biopsy needle with the same specification as that of the guiding block
into the hole of the guiding block.