Probes and Biopsy
18-37
Plastic needle-guided bracket:
(1) Put on the sterile probe sheath.
(2) Hold the probe by one hand, select the proper needle-guided bracket, and hold it
with the other hand. Align the narrow end tab of the needle-guided bracket
with the groove of the probe, then push the needle-guided bracket forward,
making the tabs and the grooves of the needle-guided bracket to match with
the grooves and tabs of the probe.
(3) Check manually to confirm that the needle-guided bracket is securely installed
on the probe.
(4) Select a proper guiding block and push it into the groove above the angle block,
and clamp it tightly.
(5) Insert a biopsy needle with the same specification as that of the guiding block
into the hole of the guiding block.
NGB-008
Metal/needle-detachable needle-guided bracket;
(1) Put on the transducer cover.