EasyManuals Logo

Mindray DC-3 User Manual

Mindray DC-3
307 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #282 background imageLoading...
Page #282 background image
DICOM
A-10
8 Scheduled Station: To set the AE Title of the scheduled station. For example, if the
AE Title of the Worklist server is M, then the scheduled station AE Title here should
be M.
9 Add: Click to add the Worklist service to the service list.
10 Cancel: Click to cancel parameter setting.
11 Update: Select an item in the service list, change the parameters in the above area,
and click [Update] to update the item in the service list.
12 Delete: Click to delete the selected service in the service list.
13 Default: Select an item in the service list, click [Default] and you can see “Y” in the
Default column.
14 Verify: Click to verify if the two DICOM application entities are normally connected.
15 Exit: Click to exit the screen.
Tips:
z In terms of “Scheduled Station AE Title”, if you set this item in the Worklist server,
then “Scheduled Station AE Title” configured here should be consistent with the
one set in the server.
For example, when you set “Scheduled Station AE Title” in the Worklist server as
A (this means you assigned the scheduled exam to A), then both “Scheduled
Station AE Title” and “AE Title” in the ultrasound system configuration should be
A.
MPPS Setting
1. Enter the “MPPS” page: “[Setup][Network Preset][DICOM Service Setting]
[MPPS].
2. Select device in the drop-down list, the available devices are DICOM services already
set in the system, enter the information.
z Click [Add] to add the service to the Service list.
z Select an item in the service list, change the parameters in the above area, and
click [Update] to update the item in the service list; click [Cancel] to cancel
changing.
z Select an item in the service list, and click [Delete] to delete the service.
z Select an item in the service list, and click [Default] to set the server to be the
default service.
3. Select an item in the service list, and click [Verify] to verify the connection.
4. Click [Exit] to confirm the setting and exit the page.