12-12 DICOM
DICOM MPPS setting items are described as follows:
Name Description
Device
After you set the server(s) in DICOM Server Setting, the name(s) will
appear in the drop-down list, select the name of the MPPS server.
Service
Name
Default is xxx-MPPS, and it can be modified.
AE Title
Application Entity title, here, it should be consistent with that of the MPPS
server.
Port
DICOM communication port, 104 by default. Here, the port should be
consistent with that of the MPPS server.
Maximum
Retries
Reserved feature.
Interval
Time(s)
Reserved feature.
Timeout
Value: 560s, in increments of 5s, and 15 by default.
Add Add the DICOM service to the service list.
Cancel Click to cancel the parameter setting.
Update
Select an item in the service list, change the parameters in the above
area, and click [Update] to update the item in the service list.
Delete Click to delete the selected service in the service list
Default
Select an item in the service list, click [Default] and you can see “Y” in the
Default column.