Chapter 12 RS-232C Communication
12.4 List of Serial Command
156 HORIBA
On-Line Operation Commands(1)
Format
Header, Item code, Parameter <CR><LF>
In RS-232C communication, the relative mV measurement is not available.
Command
for...
Command
Parameter
Description of
parameter
Note
Header
Comma
nd code
On-line/
off-line
C OL x 0: On-line
1: Off-line
ON ON, OFF OFF: The state
remains unchanged.
OFF ON: Proceeds to standby mode
for measurement in the main screen.
ON OFF: No screen transition
Effective when the screen is in Mea-
surement. For other cases, the error
response (malfunction) returns.
Halt
potential
hunting
C BR x x: channel number The measurement or calibration in Auto-
Hold is aborted.
In the measurement with AutoHold OFF
or in calibration, either calibration value
is forcefully obtained (in calibration), or
holds the measured value (in measure
-
ment).
In the interval memory, the error
response (malfunction) returns.
pH measure-
ment mode
specification
C PH x x: channel number pH Measurement proceeds to standby
mode for the specified channel.
mV measure-
ment mode
specification
command
C MV x x: channel number Falls into the measurement ready condi-
tion in the ABS. ORP measurement
mode in the specified channel.
Ion measure-
ment mode
specification
C IO x x: channel number Falls into the ion measurement ready
condition in the specified channel.
Conductivity
measure
-
ment mode
specification
C CO Falls into the COND measurement
ready condition.
Saline
mea-
surement
specification
C SA Falls into the salinity measurement
ready condition.
Resistivity
measure-
ment specifi-
cation
C OH Proceeds to standby mode for resistivity
measurement.