Chapter 3 pH MEASUREMENT
3.3 Changing the measurement setting condition
42 HORIBA
3.3.5 Electrode lot change
By inputting the electrode Lot, the data printout (detailed) and/or data memory will contain
the electrode Lot information.
Setting procedure
1. In the pH SETUP screen, move the
cursor to the ELECTRODE LOT
field by or key.
2. Press the ENTER key.
The ELECTRODE LOT screen
appears.
3. Select the numerical values by ,
and/or , keys, and press
the ENTER key.
Setting range: maximum 7 digits
4. Press F4 (RETURN).
The display returns to the pH SETUP
screen.