Chapter 1 Overview
1.7 Communication to Each Output
16 HORIBA
1.7 Communication to Each Output
Analog output
Use the optional exclusive cable.
For connection procedure or output configuration, refer to Chapter 13 Analog Output
in page 160.
Serial communication/printout
Used for PC communication and printout.
To output the data to the PC, connect the optional printer.
The exclusive serial cable and the printer cable are necessary.
Connection should be given with the power turned off. Connection with the power on may
damage the meter to cause a breakage.
Compact flash card (F-53,54,55)
For the detailed information on the compact flash card, refer to "11.5 Compact Flash
Setting( F-53, 54, 55)" p.144
Ethernet output (F-55)
To disconnect the cable, hold the connector part with a screwdriver and remove the
connector cap.
Do not pull the cable forcefully.
1.8 Operation of Display Value
This meter performs measurement value operation with the digit lower than that
displayed in displayed values or set values, and sometimes the following situation
may occur:
●
The inspection judgment rank may differ from the actual one when the inspection
result is the same with that specified in the regulation.
●
A minus indication may attach to zero in the displayed value, as -0.000.
●
Alarm may not be triggered even when the display value should show the alarm
set value.
●
The printout data may differ in 1 digit between the display value and the printout
value.