Chapter 2 BASIC OPERATIONS
2.5 Navigation Function
F-52, 53, 54, 55 27
2.5 Navigation Function
Pressing the NAVI key in the Measurement, CAL, or SETUP screen provides the following
navigation functions.
●
Operation guide display in the Measurement, CAL screens
●
Operation guide display in the SETUP screen
●
Pressing the NAVI key again in the NAVI screen returns to the previous screen.
●
In the NAVI screen, if the screen contains the page “Detailed,” press the Detailed
page mark and press the ENTER key. The display jumps to the page.
●
When an error occurs, the counteraction is displayed and the error indication is
deleted from the screen.
The error indications for ERR08 (Calibration Interval Error) and ERR10 (Memory Over) are
not deleted unless the cause of errors are removed. The screen only gives the counteraction
display; the actual performance of the counteraction is necessary to delete the error indica
-
tion.
To clear ERR08 (Calibration Interval Error), perform calibration reset or repeat the calibration
again.
To clear ERR10 (Memory Over), reset the memory data.