Chapter 5 Ion Measurement (F-53,55)
5.3 Measurement Using Standard Addition Method
74 HORIBA
5.3.3 Measuring Ion Concentration in Sample Addition 1 Mode
This mode is used to obtain the ion concen-
tration of the target ion species by adding a
small amount of the sample to the standard
solution for the ion species to measure once,
thus increasing the ion concentration of the
target component, and then measuring the
change in electric potential which occurs at
that time.
1. In the ION SETUP screen, press
↑ or ↓ key and select the MEA
-
SUREMENT option.
2. Press ← or → key, select the
INCREMENTAL option and press
F4 (RETURN).
The ION Measurement screen
appears.
3. Press ↑ or ↓ key and select the
SAMPLE ADD option.
4. Press ← or → key and select (1)
SINGLE.
5. Press F4 (RETURN).
The SAMPLE ADDITION 1 screen
appears.