Chapter 15 Maintenance and troubleshooting
15.2 Troubleshooting
206 HORIBA
● ERR No.11 Cell constant out of range
Definition
The cell constant is out of the range of 0.7 to
1.3. Delete data stored in the memory con
-
firming their contents.
● RR No.12 CF Write over defective
Definition
The compact flash card or its contents are abnormal.
● ERR No.13 CF over capacity
Definition
The capacity of the Compact flash memory is exceeded.
● ERR No.14 CF not inserted
Definition
The compact flash card is not correctly inserted.
Cause How to solve problem
COND electrode is at
the end of the useful
life.
Replace the electrode.
Improper standard
solution
Prepare standard solution.
Cause How to solve problem
Compact flash is not
formatted.
Format Compact flash.
I/F of Compact flash
card cannot be read.
File system is abnor
-
mal.
Prepare a new Compact flash.
Cause How to solve problem
Capacity of Compact
flash
memory is
exceeded.
Delete data.
Prepare a new Compact flash.
Cause How to solve problem
Compact flash is not
inserted.
Confirm that the Compact flash is correctly
inserted.