Chapter 15 Maintenance and troubleshooting
15.2 Troubleshooting
F-52, 53, 54, 55 213
● When the printing is not made even if the printer is connected
Check the following points:
・ Is the printer turned ON?
・ Is an error occurred in printer?
・ Is the printing paper run out or clogged?
・ When the test print is made according to the manual, is it done correctly?
● The settings of the meter differ from those registered.
● The operator SELECT OPERATOR screen opens instead of the Measure-
ment screen after the power is turned ON.
● Maintenance, Clock setting, and Security setting can not be selected or
changed.
Cause How to solve problem
ON/OFF was
pressed.
Make the operator selection again.
Cause How to solve problem
The user manage-
ment is still ON.
Newly register settings and make the operator
setting, or make the Administrator turn OFF
the user management.
Cause How to solve problem
A person other than
the Administrator has
made the operator
selection.
When the user management is ON: Ask the
Administrator for change.
When the user management is OFF: Restart
in the *GUEST* mode.
The user manage-
ment is set ON, but
the Administrator is
not registered.
Newly register the operator in the SELECT
OPERATOR screen at the startup, and also
register the Administrator.
Administrator's pass-
word was forgotten.
Reset to the settings made on the shipment
from the factory.
Check on our website (http://www.horiba-
water.com) or contact with our customer sup-
port center.