Chapter 16 Reference
16.4 Conductivity measuring
244 HORIBA
Salinity conversion
Salt concentration indicates the salt concentration of the intended seawater based on the nor-
mal seawater, converting it according to its measured conductivity and temperature.
Therefore, note that all the substances detected as conductivity is indicated as salt concentra-
tion.
For instance, even if a composition in a sample is hydrochloric acid (HCl), it is indicated as
salt concentration of seawater.

Table of Contents