Chapter 16 Reference
16.6 Default settings
248 HORIBA
16.6 Default settings
When reset to the settings on shipment from the factory, the following setting contents, per-
sonal information, and calibration data will be reset.
Resettingtothefactorysetvaluereturnsthesettingdatatoinitialvalues(exceptforlanguageand
clock).
Category
Item
○ : shows items to
store setting on opera
-
tor basis.
Default values
(Set value on
shipment from
factory)
Setting range (Setting item)
pH SETUP DISPLAY OPTION
○
0.001pH
0.001 pH, 0.01 pH
pH SETUP TEMP
COMPENSATION
○
AUTO AUTO, MANUAL
(0.0 ℃ to 100.0 ℃)
pH SETUP TEMP
COEFFICIENT
○
OFF
OFF, ON (COEF; -0.100pH to
+0.100pH, Temperature; 0.0
℃ to 100.0 ℃)
pH SETUP ELECTRODE
MODEL
○
No indication Selection formula from
attached material or freed
input (9 characters)
pH SETUP ELECTRODE LOT
○
No indication 7 digit integer
pH SETUP LIMIT
○
OFF OFF, ON
(pH0.000 to pH14.000)
pH
CALIBRATION
BUFFER
NIST NIST, USA, Custom
pH
CALIBRATION
CAL POINTS
2 1 to 5
pH
CALIBRATION
CAL INTERVAL
OFF OFF, ON (1-999 days)
pH
CALIBRATION
REPEATABILITY
OFF ON, OFF
pH
CALIBRATION
SYSTEM CHECK
NIST
BUFFER(
pH7,4
)
JIS(pH7,4,9),
NIST BUFFER(pH7,4),
SIMULATOR(X-51)
mV SETUP LIMIT
○
OFF OFF, ON
(-1999.9mV to 1999.9mV)
ION SETUP MEASUREMENT
○
AFTER CAL AFTER CAL, INCREMENTAL
ION SETUP TEMP COMPEN-
SATION
○
AUTO AUTO, MANUAL
(0.0 ℃ to 100.0 ℃)
ION SETUP ION SELECTION
○
ELECTRODE ELECTRODE, MANUAL
(-2,-1,+1,+2)
ION SETUP UNIT
○
g/L g/L ,mol/L