7.6 Checking Indication Accuracy
E5EK
7--13
Preparation
The following figure shows the required device connection.
10
9
8
7
6
5
4
3
2
1
20
19
18
17
16
15
14
13
12
11
21 22
23
DMM
STV
+
--
SOURCE
AC100-240V
(AC/DC24V )
~
Operation
Set the STV to the current value equivalent to the check value.
Preparation
The following figure shows the required device connection.
10
9
8
7
6
5
4
3
2
1
20
19
18
17
16
15
14
13
12
11
21 22
23
DMM
STV
+
--
SOURCE
AC100-240V
(AC/DC24V )
~
Operation
Set the STV to the voltage value equivalent to the check value.
F Current input
F Voltage input