The facilities provided on the page are as follows:
List of logged
events
Displays a list of logged events in a date/time sequence. Each event is
identified by an icon, date, time, title, duration, name and comments (if
present). An event is selected (highlighted) by clicking on it.
Delete
button
Deletes the selected event
Undo button
Undo the last action
Save As Text
button
Save the selected event as a text file for use in spreadsheets etc. For details
of saving a log file as text for use in Microsoft Excel™, see sub-section
5.8.1.1.
View Run
button
Opens either the Run Log page (see sub-section 5.8.2, page 5-31) or, if
either a Leak detection run (
) or MFC calibration run ( ) is selected, the
Leak Detect and MFC calibration log page (see sub-section 5.8.3, page 5-32)
with the selected log data displayed.
Filter by
Type list
A list of event types with associated checkboxes. Use this panel to select the
events to display in the Event list. A checkbox showing an ‘x’ indicates that
the associated event type will not be displayed. A checkbox showing a ‘9’
indicates that the associated event type will be displayed.
Filter by
Batch Name
field
Enter a batch name to list only logged events associated with that batch.
Filter by
time fields
and buttons
Use these controls to select events occurring in a time range to be
displayed.
Hide items
before
selection
button
Displays all events after and including the highlighted event.
Hide items
after
selection
button
Displays all events before and including the highlighted event.
Show all
items button
Displays all previously hidden events.
RKUKNKN p~îáåÖ=~=äçÖ=ÑáäÉ=~ë=íÉñí=Ñçê=ìëÉ=áå=jáÅêçëçÑí=bñÅÉä»=
NOTE: OIPT now supplies customers with comprehensive software to view/analyse PC 2000
log files. The software, LogViewer, is provided on the system PC. For full details of
LogViewer, refer to its Manual by following the shortcut on the desktop of your
system PC.
Any logged process run can be saved at text and then opened in Excel for viewing, analysing,
etc. To do this, use the following steps:
Operating Instructions