4: Repair Internal Assemblies — Rear Case Assembly
108
2 Remove cable tie and unplug the connectors and NBP tube, see Figure 74.
Figure 74 Measurement Module-2
Replacement
1 Install the Measurement Module and make sure it seats flat against the side wall of the Rear Case at
tis designated place.
2 Install the NBP tube and plug in EtCO
2
, SpO
2
, and ECG cables onto process PCA, and use the
cable tie to fasten the EtCO
2
cable, see Figure 74.
3 Install the Measurement Module Locking Bracket into the locking groove on connector block and
against inner side wall of the Rear Case, see Figure 73.
4 Close the Case. See “Closing the Case on page 149.
To Complete the Replacement
Run Performance Verification and Safety testing as described in the “Performance Verification”
chapter.