4: Repair Internal Assemblies — Main Chassis
114
3 Remove the five M 4x16 screws with washers (1 to 5)and two M 3x10 screws in black circle, see
Figure 84.
Figure 84 Main Chassis-3
4 Position the Rear Case Assembly of the Efficia DFM100 bottom side down.
5 Hold the plastic part of the Main Chassis and pull out the Main Chassis.
Figure 85 Main Chassis-4
1
3
4
5
2
Main Chassis