4: Repair Internal Assemblies — Main Chassis
144
2 Use a new cable tie to fix the Capacitor.
a Make the smooth side of the cable tie facing you.
b Insert the cable tie into the holes 1 and 2 in sequence and tighten the cable tie. See Figure 125.
c Insert the other cable tie into the holes 3 and 4, then tighten the cable tie. See Figure 125.
d Cut the cable tie ends.
Figure 125 Therapy Capacitor Installation-2
3 Install the two Therapy Capacitor spade connectors to J9 and J11 on the Therapy PCA, see
Figure 90 and Tab le 30 .
4 Install the Main PCA Module. See Assembling the Main Chassis on page 118.
5 Install the Main Chassis. See “Separating Main Chassis and Rear Case on page 113.
6 Close the case. See “Closing the Case on page 149.
To Complete the Replacement
Run Performance Verification and Safety testing as described in the “Performance Verification”
chapter.