Components Troubleshooting 3: Troubleshooting
59
In Table 17 “ECG Diagnostic with the Hardware Error Log, the Numeric IDs of the (9:nn) format
indicate problems with Leads ECG, and the ones of the (10:nn) format indicate problems with Pads
ECG.
Table 17 ECG Diagnostic with the Hardware Error Log
Numeric ID Info String Possible Cause Suggested Solution
 
ECG test is not completed in time Run Op Check


nn (n=0 -6) Communication failure Replace SOM PCA


nn (n=7 - 11)
Reinstall software
•Replace SOM PCA


xxx
 
 
Self-test result is not received



xxx Leads / Pads calibration failure
 
Leads / Pads test failure


xxx
 
ECG Leads gain failure Replace Processor PCA


ECG Leads clock failure


ECG Leads voltage failure


n,  xxx ECG leads failure


n,  xxx


n, xxx ECG leads failure Replace leads cable
Replace Measurement
Module
Replace Processor PCA


nn
ECG circuit failure Replace Processor PCA


<reason>
 nn
Replace Leads Cable
Replace Connect Block
Replace Processor PCA




Op Check performed without ECG
cable
Run Op Check
 
ECG Pads gain failure Replace Therapy PCA


ECG Pads clock failure


ECG Pads reference error


n, xxx ECG pads failure


n, xxx


n, xxx ECG pads failure Replace Therapy cable
•Replace Therapy Port
Replace Therapy PCA