EasyManua.ls Logo

Philips Efficia DFM100 - Page 71

Philips Efficia DFM100
220 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Components Troubleshooting 3: Troubleshooting
61
The Capacitor
charges too slowly
The device is being operated with
AC power (no battery) or the
battery power is low
Install a fully charged battery
Battery not fully charged, or
defective
Therapy PCA defective Replace Therapy PCA



message
Unable to charge to selected
energy
Unable to determine energy
•Run OpCheck
Check Hardware Error Log
Replace Therapy PCA
Table 18 Defibrillation Charging Problems (Continued)
Symptom Possible Cause Suggested Solution
Table 19 Defibrillation Discharging Problems
Symptom Possible Cause Suggested Solution
Op Check
message
During Op Check the 
menu pressed instead of the Shock button
Run Op Check and be sure to press the
Shock button
Run Controls Test to confirm
Check buttons for mechanical operation
During Op Check the device disarmed
before the user pressed the Shock button
Does not shock in Manual Defib
mode using Shock buttons on
paddles
Paddles failure Confirm paddles problem by connecting
Pads and attempting to discharge device
using Shock button on the Efficia
DFM100. Replace paddles if needed.
Therapy Port defective Replace Therapy Port
Processor PCA failure Replace Processor PCA
Does not shock in Manual Defib
or AED mode using Shock
button on Efficia DFM100
Front panel button failure Run Controls Test to see if the button is
operating. Replace UI PCA if needed.
UI PCA failure Replace UI PCA
Does not deliver correct energy
into defibrillator analyzer or
delivers no energy at all.

message
Therapy PCA failure Replace Therapy PCA
Charges OK, but aborts the
shock when press Shock or
paddle buttons
Patient impedance sensed as too high or
too low during energy delivery due to:
Pads/paddles losing contact with
patient
Pads/paddles failure
Therapy cable failure
Press the paddles firmly and check the
PCI LEDs
Replace paddles, pads, or therapy cable,
as needed
Therapy PCA failure Replace Therapy PCA

Table of Contents

Other manuals for Philips Efficia DFM100

Related product manuals