3: Troubleshooting Components Troubleshooting
62
Charges OK, but disarms
spontaneously
Device sensed Pads Off or Cable Off due
to pads losing contact with patient
Observe screen messages
Make sure loss of contact is not due to
patient movement
Pads / cable failure Replace pads or pads cable
Therapy PCA failure Replace Therapy PCA
Processor PCA failure Replace Processor PCA
 message Internal Resistors failure Replace Internal Resistors
Therapy PCA failure Replace Therapy PCA
 message An overload is detected in the paddles
circuitry
Replace paddles
Replace Therapy Port
Replace Processor PCA
Table 19 Defibrillation Discharging Problems (Continued)
Symptom Possible Cause Suggested Solution
Table 20 Defibrillation Diagnostic with the Hardware Error Log
Numeric ID Info String Possible Cause Suggested Solution


Test passed after a failure None, not an error


Test is not completed in time Run Op Check and be sure
to press the Charge button






xxx  Data transmission error Reinstall software
Replace Processor PCA
 
Therapy PCA failure Replace Therapy PCA
Replace Processor PCA


xxx  Test is not completed in time Run Op Check and be sure
to press the Charge button
 
Data transmission error Run Op Check
Replace Therapy PCA


xxx

xxx
Software versions mismatch Reinstall software


xxx Software failure
 

Therapy PCA failure
Replace the Battery
Replace Therapy PCA
 




nn xxx •Run OpCheck
Replace Therapy PCA
 

Corrupted software upgrade
file
Reinstall software
 