Components Troubleshooting 3: Troubleshooting
63
xx
Therapy Cable or connection
failure
•Run OpCheck
• Replace Therapy Cable
•Replace Therapy Port
• Replace Processor PCA
• Replace Therapy PCA
xxx Therapy PCA software
upgrade failure
• Reinstall software
• Replace Therapy PCA
Battery failure Replace the Battery
Therapy PCA failure Replace Therapy PCA
Therapy PCA failure Replace Therapy PCA
xxx
nn
Therapy Cable not detected • Replace Test Plug/Load
• Replace Therapy Cable
•Replace Therapy Port
• Replace Therapy PCA
Shock test failed • Run Op Check
• Replace Test Plug/Load
• Replace Therapy Cable
• Replace Therapy PCA
Failure to reach the requested
charge level in time
Failure to fully discharge Replace Therapy PCA
Therapy PCA failure
Autotest failed
Therapy Defib test failed Rerun Op Check,
Replace Therapy PCA
“Proceed As Is” soft key
pressed during Op Check
Rerun Op Check
SOM PCA failure Replace SOM PCA
Therapy RFU or Opcheck
test failed
Replace SOM PCA
Table 20 Defibrillation Diagnostic with the Hardware Error Log (Continued)
Numeric ID Info String Possible Cause Suggested Solution