EasyManua.ls Logo

Philips IntelliVue MX800 - To Change the Size of an Individual ECG Wave; 1 Select the wave segment you want to change. This calls up the lead menu for this segment.; 2 In the lead menu, select Size Up to increase wave size or Size Down to decrease the size.; To Change the Size of all the ECG Waves

Philips IntelliVue MX800
460 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 ECG, Arrhythmia, ST and QT Monitoring
113
Changing the adjustment factor only changes the visual appearance of the ECG wave on the screen. It
does not affect the ECG signal analyzed by the monitor.
Comparing the wave size to the 1 mV calibration bar on the ECG wave segment can help you to get an
idea of the true ECG signal strength. If you choose a fixed adjustment factor, the 1 mV calibration bar
will be the same size for all the displayed ECG waves. If you choose
Auto Size, or an individual size
using the
Size Up/Size Down keys, the calibration bar may be a different size for each wave.
To Change the Size of an Individual ECG Wave
1 Select the wave segment you want to change. This calls up the lead menu for this segment.
2 In the lead menu, select Size Up to increase wave size or Size Down to decrease the size.
Selecting
Auto Size lets the monitor choose the optimal adjustment factor for all available ECG
waves.
To Change the Size of all the ECG Waves
To change the size of all the ECG waves on the screen by a fixed adjustment factor,
1 In the Setup ECG menu, select Adjust Size.
2 Select the required adjustment factor from the line of pop-up keys.
Size X0.5 to halve the wave size
Size X1 to display the wave without zoom
Size X2 to double the wave size
Size X4 to multiply the wave size by four
Previous Size: to return one step to the previous size
Auto Size: to let the monitor choose the optimal adjustment factor for all the ECG waves.
Changing the Volume of the QRS Tone
The QRS tone is derived from either the HR or Pulse, depending on which is currently selected as the
alarm source. The QRS volume can be set from 0 to 10 (0 means off).
To change the QRS volume, in the
Setup ECG menu select QRS Volume and then select the
appropriate volume from the pop-up list.
Changing the ECG Filter Settings
The ECG filter setting defines how ECG waves are smoothed. An abbreviation indicating the filter
type is shown underneath the lead label on the monitor display. Filter settings do not affect ST
measurement.
To change the filter setting, in the
Setup ECG menu, select Filter and then select the appropriate
setting.
Monitor: Use under normal measurement conditions.
Ext. Monitor: Use when diagnostic quality is required but low frequency interference or a
wandering baseline may be expected. The upper edge frequency is the same as the
Diag setting
and the lower edge frequency is the same as the
Monitor setting.
Filter: The filter reduces interference to the signal. It should be used if the signal is distorted
by high frequency or low frequency interference. High frequency interference usually results in
large amplitude spikes making the ECG signal look irregular. Low frequency interference

Table of Contents

Related product manuals