CO2 INOPs
<Press Label>
ZERO+CHECK CAL
Numeric is replaced by -?-
Perform a zero and check the calibration of the transducer.
RAP INOPs See <Press Label> INOPs
UAP INOPs See <Press Label> INOPs
UVP INOPs See <Press Label> INOPs
INOP Message, Indication What to do
INOP Message, Indication What to do
!! CO₂ OCCLUSION
!!! CO₂ OCCLUSION
Numeric is replaced by -?-
INOP tone
The sample line or exhaust tube is blocked. Check the tubing, then disconnect and reconnect the
sample line. If the INOP persists, connect a new sample line.
CO₂ AUTO ZERO
Numeric is replaced by -?-
if the Autozero lasts >15 sec,
INOP tone sounds.
The automatic zero calibration is in progress. This typically takes 10 seconds. During this time the CO
2
values may not be updated, or they may be replaced by -?-. Wait until the zero calibration is complete
to resume monitoring.
CO₂ CAL MODE
CO
2
numeric displays current
CO
2
value for accuracy check
Currently no calibration is running. Accuracy can be checked by placing the transducer on the two cells
of the calstick and starting calibration. To start monitoring, leave Cal. Mode.
CO
2
CAL RUNNING
Numeric is replaced by -?-
Wait until calibration is finished.
CO₂ CHECK CAL
Numeric is replaced by -?-
INOP tone
The CO
2
value is outside the measurement range. Perform an accuracy check for both calstick cells
and, if necessary, recalibrate the transducer.
CO₂ CHK ADAPTER
Numeric is replaced by -?-
INOP tone
Check that the sensor is connected to the airway adapter, clean the airway adapter, if necessary.
Perform a zero calibration. If the INOP persists, contact your service personnel.
CO₂ DEACTIVATED
INOP tone
The CO
2
measurement label in the measurement device has been deactivated by deactivating the label
in the Measurement Selection window. The measurement automatically disappears from the
display. To switch the measurement on again, reactivate the measurement label in the Measurement
Selection window.
CO₂ EQUIP MALF
Numeric is replaced by -?-
INOP tone
The Measurement Extension is faulty. Unplug and replug the Multi-Measurement Module with
Extension. If you are using the mainstream method, unplug and replug the transducer or try another
transducer. If the INOP persists, contact your service personnel.
CO₂ FAILED CAL
Numeric is replaced by -?-
INOP tone
Make sure that the Cal cell was changed between CAL1 and CAL2. Repeat the calibration. If the
INOP reappears, try another transducer. If the INOP persists, contact your service personnel.
CO₂ NO SENSOR
from M3014A
Numeric is replaced by -?-
INOP tone
There is no CO
2
sensor connected. If you silence this INOP the CO
2
measurement will be switched
off.
CO₂ NO TRANSDUC
from mainstream CO
2
(except
M3014A)
Numeric is replaced by -?-
INOP tone
There is no CO
2
transducer connected. If you replace the transducer, the new transducer must be
calibrated. If you silence this INOP the CO
2
measurement will be switched off.