2Alarms
60
Alarm Latching Behavior
All INOPs are non-latching. See “Yellow Arrhythmia Alarms” on page 134 for information on one-
star yellow alarms latching behavior.
Testing Alarms
When you switch the monitor on, a selftest is started. You must check that the alarms lamps light, one
after the other, and that you hear a single tone. This indicates that the visible and audible alarm
indicators are functioning correctly. For further testing of individual measurement alarms, perform the
measurement on yourself (for example SpO
2
or CO
2
) or use a simulator. Adjust alarm limits and check
that appropriate alarm behavior is observed.
Alarm Behavior at Power On
If the monitor is switched off for longer than one minute and then switched on again (or after a loss of
power lasting longer than one minute, or when a patient is discharged), the monitor can be configured
to:
restore the alarm settings from the monitor's configured default Profile,
restore the most recently used alarm settings, or
switch the alarms off.
After any of these situations, you should check that the alarm settings are appropriate for your patient
and monitoring situation, and if necessary, select the correct Profile and patient category.
If power is lost for less than one minute, the alarm on/off condition prior to the power loss is
restored.
Red & Yellow Measurement
Alarms
Non-latching alarms Visual and audible
latching
Visual latching,
audible non-latching
Alarm has not
been
acknowledged.
Alarm condition
still present.
Alarm tone on. Alarm
lamp on. Alarm
message. Flashing
numerics.
Alarm tone on. Alarm
lamp on. Alarm
message. Flashing
numerics.
Alarm tone on. Alarm
lamp on. Alarm
message. Flashing
numerics.
Alarm condition
no longer present.
All audible and visual
alarm indicators
automatically stop.
Alarm tone on.
Alarm lamp on.
Alarm message.
Flashing numerics.
Alarm message.
Flashing numerics.
Audible alarm indicators
automatically stop.
Alarm has been
acknowledged.
Alarm condition
still present.
Alarm tone off. Alarm
lamp off. Alarm
message. Flashing
numerics. Audible alarm
reminder (if configured).
Alarm tone off. Alarm
lamp off. Alarm
message. Flashing
numerics. Audible alarm
reminder (if configured).
Alarm tone off. Alarm
lamp off. Alarm
message. Flashing
numerics. Audible alarm
reminder (if configured).
Alarm condition
no longer present.
Audible and visual alarm
indicators automatically
stop.
Audible and visual alarm
indicators automatically
stop.
Audible and visual alarm
indicators automatically
stop.