3 Patient Alarms and INOPs
76
Temperature INOPs
NBP DEACTIVATED
INOP tone
The NBP measurement label in the measurement device has been deactivated by deactivating the label
in the Measurement Selection window. The measurement automatically disappears from the
display. To switch the measurement on again, reactivate the measurement label in the Measurement
Selection window.
NBP EQUIP MALF
Numeric is replaced by -?-
INOP tone
Remove the cuff from the patient. The NBP hardware is faulty. Contact your service personnel.
You can silence this INOP, but the INOP message remains visible until the next measurement is
started or the Stop All SmartKey is selected.
NBP INTERRUPTED
Numeric is replaced by -?-
INOP tone
Check the tubing and cuff for leakages or kinks. Check that you are using the correct cuff size and
placement, and that the correct patient category is selected. Try restarting the measurement.
If the INOP occurs repeatedly, contact your service personnel.
You can silence this INOP, but the INOP message remains visible until the next measurement is
started or the Stop All SmartKey is selected.
This INOP arises when the measurement needed longer than the maximum time for inflation,
deflation or the total measurement.
NBP MEASURE FAILED
Numeric may be displayed with a
-?-
INOP tone
Check that you are using the correct cuff size and placement, and that the correct patient category is
selected. Try restarting the measurement.
You can silence this INOP, but the INOP message remains visible until the next measurement is
started or the Stop All SmartKey is selected.
Check the condition and suitability of the patient for NBP monitoring. Use another cuff to continue
measuring.
INOP Message, Indication What to do
INOP Message, Indication What to do
T1, T2, T3, T4 INOPs See <Temp Label> INOPs
Tamb INOPs See <Temp Label> INOPs
Tart INOPs See <Temp Label> INOPs
Tcereb INOPs See <Temp Label> INOPs
Tcore INOPs See <Temp Label> INOPs
<Temp Label>
DEACTIVATED
INOP tone
A Temp measurement label in the measurement device has been deactivated, either by connecting a
Pressure transducer in the shared Press/Temp socket, or by deactivating the label in the
Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either reconnect a Temp transducer or reactivate the
measurement label in the Measurement Selection window.
<Temp Label> EQUIP
MALF
Numeric is replaced by -?-
INOP tone
Contact your service personnel.
The temperature hardware is faulty.
<Temp Label> NO
TRANSDUC
Numeric is replaced by -?-
INOP tone
Make sure the TEMP probe is connected to the MMS or module.
If you silence this INOP, the measurement will be switched off.
<Temp Label>
OVERRANGE
Numeric is replaced by -?-
INOP tone
Try changing the application site of the transducer.
[The temperature is less than -1°C, or greater than 45°C.]