Installation Safety Information
WARNING
If multiple instruments are connected to a patient, the sum of the leakage currents may exceed the
limits given in IEC/EN60601-1, IEC60601-1-1, UL60601-1. Consult your service personnel.
Connectors
The actual placement of boards and configuration of connections for your monitor depends on how
your hardware has been configured. See the symbols table (“Symbols” on page 391) to see which
symbols are used to mark the connections.
WARNING
Connect only medical devices to the ECG output connector socket.
Masimo SET technology Nellcor OxiMax
compatible
Uses FAST SpO
2
algorithm
Silence Alarms Alarms On/Off Alarms Switched Off
Main Screen SmartKeys Back key
ECG Sync Output/
Analog ECG Output
Symbols
Grounding The monitor must be grounded during operation. If a three-wire receptacle is not
available, consult the hospital electrician. Never use a three-wire to two-wire adapter.
Equipotential
Grounding
If the monitor is used in internal examinations on the heart or brain, ensure that the room
incorporates an equipotential grounding system to which the monitor has a separate
connections.
Combining equipment Combinations of medical equipment with non-medical equipment must comply with IEC
60601-1-1. Never use a multiple portable socket-outlet or extension cord when
combining equipment unless the socket outlet is supplied specifically for use with that
equipment.
Fusing The monitor uses double fusing (line and neutral).
Network Cables All network cables must be unshielded.