Interfering Gas and Vapor Effects On CO2 Measurement Values
The specified deviations for M3002A/M3014A are valid when the appropriate corrections are switched on and set
correctly.
CO
2
Alarm
Specifications
Range Adjustment Delay
etCO
2
High 20 to 95 mmHg (2 to 13 kPa) 1 mmHg (0.1 kPa) M3002A/M3014A/M3016A: less than
14 seconds
M3015A: less than 21 seconds.
etCO
2
Low 10 to 90 mmHg (1 to 12 kPa)
imCO
2
High 2 to 20 mmHg
(0.3 to 3.0 kPa)
steps of 1 mmHg
(0.1 kPa)
M3002A/M3014A/M3016A: less than
14 seconds
M3015A: less than 21 seconds.
awRR High Adult/pedi: 10 to 100 rpm
Neo: 30 to 150 rpm
under 20 rpm: 1 rpm
steps
over 20 rpm:5 rpm
steps
M3002A/M3014A/M3016A: less than
14 seconds
M3015A: less than 21 seconds.
awRR Low Adult/pedi: 0 to 95 rpm
Neo: 0 to 145 rpm
M3015A:
settings <20 rpm: less than 8 seconds
>20 rpm: less than 21 seconds
M3002A/M3014A/M3016A
settings <20 rpm: less than 4 seconds
>20 rpm: less than 14 seconds
Apnea delay 10 to 40 seconds 5 second steps set apnea delay time + 4 seconds
(M3002A/M3014A/M3016A) or
8 seconds (M3015A with 2 m sample
lines) or 11 seconds (M3015A with 4 m
sample lines).
Gas or Vapor
Gas Level
(% volume fraction)
M3002A/M3014A M3015A
Additional deviation due to gas interference,
measured at 0 - 40 mmHg CO
2
Nitrous Oxide 60% ±1 mmHg ±5%
Halothane 4% ±2 mmHg ±5%
Enflurane 5% ±2 mmHg ±5%
Isoflurane 5% ±2 mmHg ±5%
Sevoflurane 5% ±2 mmHg ±5%
Xenon 80% -5 mmHg ±5%
Helium 50% ±1 mmHg ±5%
Metered dose inhaler
propellants
- not specified for use
Desflurane 15% +5 mmHg ±5%
Ethanol 0.1% ±1 mmHg ±5%
Isopropanol 0.1% ±1 mmHg ±5%
Acetone 0.1% ±1 mmHg ±5%
Methane 1.0% ±1 mmHg ±5%